MNL-1029 Rev A
20
Since the integration is not predefined (user defined), the minimum integration time is no longer 1 µs. The minimum integration
effectively becomes the maximum busy time 1, t
BUSY1
, which is 218 µs. The user can still define an acquisition delay; however, the
minimum integration remains fixed. External Level Trigger timing parameters are described below.
Start of Acquisition Cycle
External Level Trigger Pulse Width
30 ns + 218 µs
(Minimum, t
ACQDLY
= 0)
External Edge Trigger Delay
User specified, 1 µs resolution
Integration period is equal to t
LPUL
–
t
ACQDLY
– 30 ns.
User specified, 1 µs resolution
User specified, 1 µs resolution
User specified, 1 µs resolution
Continuous Strobe Off Time
The following timing diagram represents the acquisition process for an External Level trigger event.