Conditioning Quick Guide: BIOM® 5 G/55185/0818/en
Functional test: Check
Smooth movement of the articulated joint
Function of the safety rod of the front loupe. After conducting this check, the control mark must be at position (1)
Function of the focussing knob
whether all fastening screws are present
Transport Pack the products or the sterilisation tray with the components in accordance with standards DIN EN 868 / ANSI AAMI ISO
11607
Sterilising Fractionated Pre-Vacuum Process
Material: Validated steam steriliser, in acc. with DIN EN
13060 or DIN EN 285 and DIN EN 17665-1/GOST R ISO
11134
Parameters:
3 Pre-vacuum phases
Sterilisation temperature: 132°C
Minimum exposure time: 3 min
Drying time: 1 min
Gravitation Process:
Minimum Parameters:
Sterilisation temperature: 132°C
Minimum exposure time: 15 min
Drying time: 1 min
The maximum sterilisation temperature is 134°
(plus tolerance in acc. with GOST R ISO 11134.
Devices and substances validated in
studies
Laboratory Washer: Miele G 7735 CD Detergent: neodisher MediClean (concentration 0.5%,
Dr Weigert, Hamburg