1.2
SPECIFICATIONS
Size:
Lamp
and
arm
fully
extended
-
34
inches
long
Lamp
housing
and
yoke
12
inches
long
by
6-1/2
inches
wide
by
4-1/2 inches high
Weight:
Rail/Wall
Mount
Model
-
approximately
24
pounds
unpacked
Floor Stand Model - approximately 37 pounds unpacked
Voltage: 115
volts
60 Hz
Current:
8.5
amp
Fuse: 3 amp, 125 volts, 1-1/4" slo-blow
Lamp type: Halogen, 150
watt,
20
Volt,
type Burton
Irradiance: 3 to 76 microwatts/square centimeter/nanometer in the
425-475 nanometer spectrum at bed level, depending on
lamp to patient distance.
Lamp intensity: 6,500 foot candles
Lamp
spot
size:
Variable
from
3 to
20
inches
depending
on
aperture
size selected and light to mattress distance.
Lamp
color
temperature:
3,150
degrees
Kelvin
Average
lamp
life
rating:
1,800
hours
(lamp
life
will
vary
between
1000 and 3000 hours typically)
Leakage
current:
With
the
ground
wire
open
.less
than
100
micro
amperes. * >
With
the
ground
wire
closed
less
than
5
micro
amperes. *
* Measured at any exposed metal surface on
115
volt
units.
page 3