Press and release either the or button. The control
solution bottle
( ) will appear, then press .
check the expiration date before performing a control
solution test. Do not use if expired. Please write down the
expiration date on the control solution bottle after rst
opening.
Gently shake the control solution before use.
Blood Glucose Test Strips
125
Firmly insert the test strip into the meter test port. Insert
rmly in the direction of the arrow on the test strip (arrow
up). Do not insert the glucose test strip upside down.
Discard the 1
st
drop of control solution, this will eliminate
any residue.
Place a drop of control solution on a clean, dry surface
(e.g. the lid of the test strip vial).
Dip the test strip into the control solution.
Results will appear in 5 seconds. Compare the result to the
range printed on the test strip vial, the result should fall
within that range.
CAUTION
The control solution bottle icon ( ) contains the letter ‘c’ so it will not to be confused with the
medication bottle icon. This will allow you to dierentiate between a control solution test and an
actual blood test.