OKRA OKRA
68
69
OKRA Pro test strips have the new bio-sensor technology, requiring only a tiny drop
of blood. OKRA Pro meter can store 365 readings in memory with date & time. All your
results are easily downloadable to your computer.
Store the test strip vial in a cool, dry place between 36-86°F (2-30°C). Keep out of
direct sunlight. Do not freeze.
Store test strips in its original vial only. Do not mix the test strips in new vials or in any
other container.
Immediately replace the cap and close tightly after removing a test strip.
• Use test strip immediately after removing it from the vial.
• Do not use test strips after the expiration date. It may cause inaccurate results.
Make a notation of the discard date when you rst open it.
Discard any remaining test strips after 6 months of rst opening.
Avoid getting dirt, food, and water on the test strip. Do not handle test strips with
wet hands.
• Do not bend, cut, or alter test strips in anyway.
General
Storage and Handling
Cautions and Safety Information section
• For in vitro diagnostic use only
The single-patient use system is for single-patient use only and should not be shared.
• Not for neonatal use
• Do not use for diagnosis of or screening for diabetes mellitus.
Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar
patients
Alternative site testing (AST) should only be performed during periods of steady-state blood
glucose conditions (when glucose is not changing rapidly).
AST should not be used to calibrate continuous glucose monitors (CGMs).
• AST should not be used for insulin dose calculations.
All parts of the kit are considered biohazardous and can potentially transmit infectious diseases,
even after you have performed cleaning and disinfecting.
• Always use a new, sterile lancet. Lancets are for single use only.
Avoid getting hand lotions, oils, dirt or debris in or on the lancets and the reusable landing
device.
Please refer to the Instructions for use for cleaning and disinfecting OKRA Pro glucose monitor-
ing system.
Make sure to ll the strip with sucient amount of blood to avoid inaccurate result.
Verify the meter code matches the code number printed on the test strip vial.