Table 9. Data analysis at THIGH
Table 10. Regressions analysis at THIGH
Patient THIGH vs YSI 2300 Finger
Patient THIGH vs YSI 2300 Finger
System accuracy results for glucose concentration<75 mg/dL (4.2 mmol/L)
System accuracy results for glucose concentration≥75 mg/dL (4.2 mmol/L)
Within ± 5mg/dL
(within ± 0.28 mmol/L)
Within ± 10 mg/dL
(within ± 0.56 mmol/L)
Within ± 15mg/dL
(within ± 0.83 mmol/L)
1/4 (25%) 3/4 (75%) 4/4 (100%)
Within ±5% Within ±10% Within ±15% Within ±20%
98/146 (67%) 124/146 (85%) 140/146 (96%) 146/146 (100%)
Linear regression 95% CI Slope 95% CI Intercept R
2
N
y = 0.9827x + 1.3423 (0.965, 1.001) (-3.693, 6.377) 0.9875 150
Table 11. Data analysis at CALF
Table 12. Regressions analysis at CALF
Patient CALF vs YSI 2300 Finger
Patient CALF vs YSI 2300 Finger
System accuracy results for glucose concentration<75 mg/dL (4.2 mmol/L)
System accuracy results for glucose concentration≥75 mg/dL (4.2 mmol/L)
Within ± 5mg/dL
(within ± 0.28 mmol/L)
Within ± 10 mg/dL
(within ± 0.56 mmol/L)
Within ± 15mg/dL
(within ± 0.83 mmol/L)
3/4 (75%) 4/4 (100%) 4/4 (100%)
Within ±5% Within ±10% Within ±15% Within ±20%
101/146 (69%) 128/146 (88%) 140/146 (96%) 146/146 (100%)
Linear regression 95% CI Slope 95% CI Intercept R
2
N
y = 0.9907x + 5.6915 (0.974, 1.008) (0.878, 10.505) 0.9888 150