4.3 Using EndoTherapy accessories
BF-190 Series OPERATION MANUAL
Ch.4
WARNING
Do not bring the distal end of the APC probe close to a metal stent. The tissue
around the metal stent may be burned.
Set the APC unit to the minimum necessary output level. If the output level is too
high, the endoscope’s and/or accessory’s insulation may be damaged and cause
patient and/or operator burns.
Before performing APC, inspect the surface of the endoscope for any dents, bulges
or other irregularities. The endoscope may be damaged and cause patient and/or
operator burns.
NOTE
The outer surfaces of the BF-190 series endoscopes except the BF-XP190 are
insulated. This allows APC to be performed.
Some Olympus endoscopes are equipped with a feedback circuit to lead leakage
current from the endoscope to the electrosurgical unit. However, the BF-190 series
endoscopes are not equipped with a feedback circuit because leakage current from
the electrosurgical accessory to the endoscope is minimal as the insertion tube is
short. Therefore, the S-cord is unnecessary when using the BF-190 series
endoscopes.
Prepare, inspect, and connect the electrosurgical unit, APC unit, and electrosurgical
accessories as described in their instruction manuals.