54
Chapter 5 Troubleshooting
BF TYPE 160 Series OPERATION MANUAL
Endo-therapy accessories
Others
Irregularity
description
Possible cause Solution
An endo-therapy
accessory does not
pass through the
instrument channel
smoothly.
An incompatible
endo-therapy accessory
is being used.
Refer to the “System chart” in the
Appendix and select a compatible
endo-therapy accessory. Confirm
that the color code on the
endo-therapy accessory matches
that on the endoscope.
Irregularity
description
Possible cause Solution
The remote switch does
not work.
The wrong remote
switch is operated.
Operate the correct remote
switch.
The remote switch
function has been set
improperly.
Set the remote switch function
correctly as described in the video
system center’s instruction
manual.