Important Information — Please Read Before Use
7
BF TYPE 160 Series OPERATION MANUAL
Do not pull the universal cord during an examination. The
endoscope connector will be pulled out from the output
socket of the light source and the endoscopic image will not
be visible.
Do not coil the insertion tube or universal cord into a diameter
of less than 12 cm. Equipment damage can result.
Do not apply shock to the distal end of the insertion tube,
particularly the objective lens surface at the distal end. Visual
abnormalities may result.
Do not twist or bend the bending section with your hands.
Equipment damage may result.
Do not squeeze the bending section forcefully. The covering
of the bending section may stretch or break and cause water
leaks.
The endoscope’s remote switches cannot be removed from
the control section. Pressing, pulling or twisting them with
excessive force can break the switches and/or may cause
water leaks.
Do not attempt to bend the endoscope’s insertion tube with
excessive force. Otherwise the insertion tube may be
damaged.
The endoscope contains a memory chip that stores
information about the endoscope and communicates this
information to the CV-160. Although the memory chip is
durable, damage will prevent data from being backed up on
it. When data are lost or damaged, contact Olympus.