Chapter 4 Operation
73
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
When the tip of the endo-therapy accessory extends
approximately 1 cm from the distal end of the endoscope, the
accessory will appear in the endoscopic image.
When the accessory appears in the endoscopic image, it
may also reflect the light from the endoscope and/or cast its
shadow in the endoscopic image. This does not indicate a
malfunction (for GIF-N180 only).
Operation of endo-therapy accessories
Operate the endo-therapy accessory according to the directions given in its
instruction manual.
Withdrawal of endo-therapy accessories
Patient debris might spray when the endo-therapy
accessories are withdrawn from the biopsy valve. To prevent
this, hold a piece of gauze around the accessory and the
biopsy valve during withdrawal.
Do not withdraw the endo-therapy accessory if the tip is open
or extended from its sheath; patient injury, bleeding,
perforation and/or instrument damage may occur.
Withdraw the endo-therapy accessory slowly and straight out
of the biopsy valve. Otherwise, the valve’s slit and/or hole
could be damaged. This can reduce the efficacy of the
endoscope’s suction system, and may leak or spray patient
debris or fluids, posing an infection-control risk.
If the endo-therapy accessory cannot be withdrawn from the
endoscope, close the endo-therapy accessory and/or retract
it into its sheath, then carefully withdraw both the endoscope
and the endo-therapy accessory together under endoscopic
observation. Take care not to cause tissue trauma.
Withdraw the endo-therapy accessory slowly while the tip of the endo-therapy
accessory is closed and/or retracted into its sheath.