U-RFL-T
6
TROUBLESHOOTING GUIDE
a. The main switch LED does not The power cord is not connected. Connect the power cord correctly.
light.
b. Main switch indicator lights, Improperly connected connectors. Connect correctly.
but the mercury burner
The burner is not installed. Install the burner.
does not ignite.
The lamp housing interlock Tighten the burner socket clamping screw.
mechanism is activated.
Auto ignition is malfunctioning. Turn the main switch on the power supply
to “O" (OFF). Turn to “I" (ON) again.
(Repeat as necessary.)
c. The burner flickers or is dark. Insufficient time has elapsed since Wait for 10 minutes after turning on the
the burner was turned on. burner.
The burner life has expired. Replace the mercury burner when the
hour counter reading exceeds 200 hours
(USH102D) or 300 hours (HBO103W/2).
Problem Cause Remedy
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If problems occur,
please review the following list and take remedial action as appropriate. If you cannot solve the problem after checking the entire
list, please contact your local Olympus representative for assistance.