Operator maintenance
NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL. REMOVING
THE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES. THERE ARE NO OPERATOR
MAINTAINABLE PARTS WITHIN THE EQUIPMENT.
In the unlikely event that you experience any problems with your unit which cannot easily be remedied, you
should contact your supplier and return the unit if necessary. Please include any details of the fault observed and
remember to return the unit in its original packing. Omega will accept no responsibility for any damage to units
that are improperly packed for shipment. If in doubt, contact your supplier. See the Decontamination Certificate
supplied with your unit.
1.
Cleaning: Before cleaning your unit, ALWAYS disconnect it from the power supply and allow it to
cool below 50° C. Your unit can be cleaned by wiping with a damp soapy cloth. Care should be
exercised to prevent water from running inside the unit. Do not use abrasive cleaners.
2.
Fuses: Your unit is protected by two fuses. They should only be changed by suitably qualified
personnel. If the fuses blow persistently, a serious fault is indicated and you may need to return the
unit to your supplier for repair.
ADDITIONAL INFORMATION
The controller is factory preset. See the Controller book for further details.
For safety reasons, the configuration settings of the controller are password protected.
Please contact your dealer for further information.
Listed below are important control parameters for the unit.
Level 1> band=11, int.t =1.5, dEr.t=12, DAC=1.0, Cyc.t=0.3, Bnd.2=15, Cyc.2=20,
Level 2 > SP2.A=cool, diSP=0.1, Hi.Sc=400.0, inpt=rtd,
Level 3 > SP1.d=SSR, SP3.d=rly
-7-
R
e
placement Parts
The following parts may be obtained from Omega Engineering if replacements or alternatives are required:
Part Number Description
FCABLEUK UK 240 volt Mains cable with 13amp UK plug (5 amp fuse)
FCABLEEU Euro style 240 volt Mains cable with R/A Schuko plug
7002705 US 120 volt Mains cable
7032722 insert extractor
7032718 Carrying case
Part Number
Available inserts
Insert well Sizes
CL701B insert 5 x 1/4”
CL702B insert 1 each 1/8”, 3/16”, 1/4”, 5/16”, 3/8”
CL703B insert 2 x 1/4” & 2 x 3/8”
CL704B insert 2 x 1/4” & 2 x 1/2”
CL705B insert 1 x 1/4”
CL706B insert blank
CL707B insert 1 x 9/16” & 1 x 1/4”
CL708B insert 1 x 5/8” & 1 x 1/4”
CL709B insert 1 x 11/16” & 1 x 1/4”
CL710B insert 1 x 3/4” & 1 x 1/4”
Part Number
Spare Parts
Description
7002695 225 watt, 120 volt heater
7002697 PRT
7002698 Solid state relay
7007201 8 amp fuse (120 volt units)
6500131 4 amp fuse (240 volt units)