EasyManua.ls Logo

Omega ILD24-UTP - Approval Information

Omega ILD24-UTP
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64
APPROVALINFORMATION
1. Electromagnetic Compatibility (EMC)
This device conforms with requirements of EMC Directive 89/336/EEC,
amended by 93/68/EEC. This instrument complies with the following EMC
Immunity Standards as tested per EN 61326-2:1998 (Industrial Locations)
Phenomena Test Specification Basic Standard
Electrostatic +/- 4 kV contact discharge IEC 1000-4-2
Discharge +/- 8 kV air discharge Performance
Criteria B
Radio Frequency 27 - 1000 MHz IEC 1000-4-3
electromagnetic 3 V/m Performance
field. 80% AM (1 KHz) Criteria A
Fast Transients +/- 2 kV (ac mains) IEC 1000-4-4
+/- 1 kV (dc, signal I/O) Performance
5/50 ns Tr/Th, 5 KHz rep. freq. Criteria B
Surge +/- 1.0 kV AC Port L-L IEC 1000-4-5
+/- 2.0 kV AC Port L-PE Performance
+/- 1.0 kV DC Port L-L Criteria B
+/- 2.0 kV DC Port L-PE
Radio Frequency 0.15 - 80 MHz IEC 1000-4-6
conducted 3 Vrms Performance
80% AM (1 KHz) Criteria A
This instrument complies with the following EMC Emission Standards as
tested per EN 61326-2:1998 (Class B equipment)
Phenomena Frequency Limits Basic
Range Standard
Radiated 30-230 MHz 30 dB_V/m at 10 m CISPR 22
Emission 230-1000 MHz 37 dB_V/m at 10 m Class B
quasi peak
Conducted 0.15-0.5 MHz 66-56 dB_V quasi peak CISPR 22
Emission 0.5-5 MHz 56 dB_V quasi peak Class B
5-30 MHz 60 dB_V quasi peak
2.Safety
This device conforms with Low Voltage Directive 73/23/EEC, amended by 93/68/EEC.
The following LVD requirements have been met to comply with EN 61010-1, 1993
(Electrical equipment for measurement, control and laboratory use)
1. Pollution Degree 2
2. Installation Category II
3. Class I Equipment (100-240 Vac Powered Units)

Table of Contents