5-7
Frequently Used Setup Operations (Setting Mode)
5
Square Root Computation
Select SQRT.
Set the channel range.
Select the range type of DC voltage.
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V,
or 50 V
Set the left span value.
Set the right span value.
Set the left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
Set the low-cut point.
0.0 to 5.0% of the recording span, 0.1 steps.
The new setting takes effect.
First channel
Last channel
ESC/?
Set=Range
CH=01-01
Mode=SQRT
Range=2V
Span_L= -2.000
Span_R= 2.000
Scale_L= 0.00
Scale_R= 200.00
Low-cut=Off
Low-cut point=0.0
01-01 Channel
Setting complete
Description
Input Type in Which Square Root Computation Can Be Performed
Square root computation can be performed when set to DC voltage input.
Span Left and Span Right
Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.
Scale Left and Scale Right
Selectable range (mantissa): –20000 to 30000
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Make sure that scale left is less than scale right.
Low-cut and Low-cut Point
If low-cut is set to On, set the low-cut point.
Selectable range of the low-cut point: 0.0 to 5.0% of the recording span
Values below the low-cut point is set to 0% (scale left value) of the recording span.