• Cleaning the Nebulizer Kit, Mask, Mouthpiece, Nasal Shower and Adapter:
Fully disassemble the Nebulizer Kit and the Nasal Shower. Wash them in warm water
and mild, neutral detergent. Rinse them thoroughly with clean hot tap water, gently
tap to remove excess water and allow to air dry in a clean place. It is advisable to
replace the Nebulizer Kit and the Nasal shower after some 100 to 120 treatments or
after about 20 boiling cycles.
Cleaning the Compressor and Air Tube:
Firstly, make sure that the power plug is unplugged from the power outlet. Wipe
clean with a soft cloth moistened with water or mild, neutral detergent.
Disinfecting the Nebulizer Kit, Mask, Nasal Shower, Mouthpiece and Adapter:
Always disinfect the parts before using the product for the first time and after the
product has not been used for an extended period of time as well as after the last
treatment of the day. If the parts are heavily stained, replace them with new ones.
The Nebulizer Kit, Mask, Nasal Shower, Mouthpiece and Adapter can be disinfected
by
using a chemical disinfectant such as ethanol, sodium hypochlorite (Milton),
Quaternary ammonium (Osvan), Chlorhexidine (Hibitane) and Amphoteric Surfactant
(Tego), following product instructions and rinsing off with clean warm water. Leave to
air-dry. (Note: Never clean with benzene, thinner or a flammable chemical.)
boiling: The Nebulizer Kit, Nasal Shower, Mouthpiece and Adapter can also be
disinfected by boiling between 15 to 20 minutes in plenty of water. After boiling,
carefully remove the parts, shake off excess water and allow to air-dry in a clean
environment. (Note: Do not boil the Mask and the Air Tube.)
Replacing the Air Filter:
If the air filter has changed colour, or has been used for more than 70 applications,
replace it with a new one. To purchase the filters see Other Optional/Replacement
Partsparagraph. Remove the air filter cover using a flat screwdriver as shown in
figure; remove the filter and fit the new filter as shown. Put the air filter cover back in
place. (Note: Do not wash or clean the air filter. If the air filter Becomes wet, replace
it. Damp air filters can cause blockages.)
Icons in the IM:
Class II equipment
Alternating current
Type BF applied part
Serial number

Other manuals for Omron C102 Total

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron C102 Total and is the answer not in the manual?

Omron C102 Total Specifications

General IconGeneral
BrandOmron
ModelC102 Total
CategoryRespiratory Product
LanguageEnglish

Related product manuals