8. Technical data
EN
Technical data for the OMRON NE-C801KD Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:
Result of cascade impactor measurements for particle size**
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009
Note: Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s
data sheet for further details.
Particle Size:
**MMAD Approx. 3µm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity:
7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Nebulization Rate:
*Approx. 0.3ml/min (by weight loss)
Aerosol Output:
**0.47ml (2ml, 1%NaF)
Aerosol Output Rate:
**0.06ml/min (2ml, 1%NaF)
0
10
20
30
40
50
60
70
80
90
100
1000110.1
Cumulative % particle mass of sodium fluoride under size
Particle Size Dp (µm)
Individual
Mean
MMAD 3.0 µm
Cumulative Undersize %
NE-C801S-KDAP(V)_A_M.book Page 23 Tuesday, July 23, 2013 12:23 PM