Filter
Check once a month whether the filter in the filter compartment is still white and clean. If it turns dirty,
replace it. However this filter should be replaced at least once a year.
Procedure for replacing:
1. Slide the filter cover open to expose the filter compartment.
2. Remove the dirty filter with a small, pointed object.
3. Replace with a clean OMRON filter (order no.8403).
A set of 5 spare filters is provided with the nebuliser.
4. Slide the filter cover closed.
Caution: Do not use cotton or any other material to replace the filter. Do not wash or clean the filter.
Use only OMRON filters. Do not operate without a filter.
Possible problems and how to solve them
If the OMRON CX3 does not nebulise, or only very weakly, check the following points:
l Make sure that the power supply plug is firmly fitted onto the wall socket.
l Make sure that the ends of the air tube are fitted tightly onto the compressor unit and onto the
medication tank.
l Check whether the medication tank is empty. If so, and if necessary, refill the medication tank.
l Is the vaporiser head installed correctly in the medication tank? If not, install it and press it down
into the jet nozzle.
l Check whether the air tube is twisted. If so, untwist.
l Check whether the jet nozzle is blocked. If so, unblock it by cleaning the medication tank.
If not possible to unblock, replace the nebuliser kit.
l Check whether the air filter is clean. If not, replace it (refer to section 'Filter' in this manual).