Troubleshooting for: main unit, nebulizer kit and
nasal aspirator
Check the following if the device should fail before or during operation.
Problem Cause Remedy
Nothing happens
when the power
switch is turned.
Is the power plug plugged
into a power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary.
No nebulization/
aspiration or low
QHEXOL]DWLRQZKHQ
the power is on.
Is there sufficient medication
in the medication tank?
RQO\QHEXOL]HUNLW
Add the correct amount of
medication to the medication
tank.
Is the vaporizer head missing
or not assembled correctly?
RQO\QHEXOL]HUNLW
Attach the vaporizer head
correctly.
Is the nebulizer kit or
nasal aspirator assembled
correctly?
Assemble the nebulizer kit or
nasal aspirator correctly.
Is the nozzle for the nebulizer
kit or aspirator blocked?
Make sure that the nozzle is free
RIEORFNDJHVGRQRWSLQFKWKH
nozzle.
Is the nebulizer kit or nasal
aspirator tilted at a sharp
angle?
Make sure that the nebulizer kit
is tilted at an angle less than
GHJUHHVRUWKHDVSLUDWRULV
upright.
Is the air tube connected
correctly?
Make sure that the air tube is
correctly connected to the
compressor and nebulizer kit or
nasal aspirator.
,VWKHDLUWXEHIROGHGEORFNHG
or damaged?
Make sure that the air tube
does not contain kinks or is not
damaged or blocked.
Is the air filter dirty? Replace the air filter with a new
one.
7KHFRPSUHVVRULV
abnormally loud.
Is the air filter cover attached
correctly?
Attach the air filter cover
correctly.
7KHFRPSUHVVRULV
very hot.
7KHFRPSUHVVRULV
insufficiently ventilated.
Do not cover the compressor
during use.
Do not block the ventilation slots.
Do not use the device in an
environment with temperatures
higher than 4C.
7KHFRPSUHVVRU
suddenly stops during
use.
7KHWKHUPDOFXWRXWKDV
shut the device down due to
overheating
See note below.
Note:
Do not attempt to repair the device. Do not open and/or tamper with the device. No parts
of the device are user serviceable. Return the device to an authorized OMRON retail
outlet or distributor.