Troubleshooting
19
Troubleshooting Troubleshooting
If any trouble occurs while you are using the unit, please check the following first.
Replace the batteries with new ones.
(Refer to page 9.)
Clean the Mesh Cap by boiling to
remove the stain. (Refer to Page 16.)
- If the trouble cannot be corrected by
the above-mentioned procedure,
replace the Mesh Cap with a new
one. (Refer to Page 17.)
Insert batteries in the correct
alignment.
(Refer to Page 9.)
Replace batteries with new ones.
(Refer to Page 9.)
Remove the AC Adapter from the
Main Unit and connect it correctly.
(Refer to Page 8.)
Recharge the batteries with a
commercially available recharger or
replace them with new ones.
Install the Medication Bottle correctly.
(Refer to Page 11.)
Fill the medication.
(Refer to Page 10.)
Clean the Mesh Cap by boiling to
remove the stain. (Refer to Page 16.)
- If the trouble cannot be corrected by
the above-mentioned procedure,
replace the Mesh Cap with a new
one. (Refer to Page 17.)
Replace the Mesh Cap with a new
one. (Refer to Page 17.)
Use the unit by slightly slanting the
Main Unit (the side of the Button)
towards yourself. (Refer to Page 13.)
Remove the medication or water.
(Refer to Page 13.)
Wipe off the stain. (Refer to Page 15.)
If the unit does not nebulize normally after taking the above-mentioned procedure, contact the store
where you purchased the unit or the nearest OMRON dealer.
Trouble
The
nebulization
rate is
extremely low.
The Power
Indicator does
not turn on and
the unit does
not nebulize.
The Power
Indicator turns
on but the unit
does not
nebulize.
Possible cause How to correct
The Mesh is stained.
The polarities (+ and -) of batteries
are in the wrong direction.
The batteries are low.
The AC Adapter is not correctly
connected to the electric outlet
of the Connection Port of Main
Unit.
The capacity of rechargeable
batteries (nickel-hydrogen
batteries) is low.
The Medication Bottle is not
installed correctly to the Main Unit.
The Medication Bottle is not
filled with medication.
The Mesh is extremely stained.
The Mesh is broken.
Medication is not supplied to the
Medication Bottle. (Medication is
not in contact with the Vibrator.)
Medication or water is collected
on the Electrodes of the Main
Unit.
The Electrode of the Main Unit
is stained.
The Battery Low Indicator
flashes / turns on, and the
batteries are low/worn out.
EN-p2-20-(NE-U22)-3 3/25/02 4:48 PM Page 19

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron Micro A-I-R NE-U22 and is the answer not in the manual?

Omron Micro A-I-R NE-U22 Specifications

Power IconPower
power source3 VDC (two AA alkaline batteries), 2.4 VDC (two AA nickel-hydrogen batteries)
power consumption1.5 W
Technical Specifications IconTechnical Specifications
ultrasonic frequency180 kHz
nebulization rate0.25 ml / minute or more
particle size5 µm
Capacity and Dimensions IconCapacity and Dimensions
medication bottle capacity7 ml
external dimensions38 (W) x 104 (H) x 51 (D) mm
weight97 g
Battery and Operating Life IconBattery and Operating Life
battery life with continuous nebulization4 hours
battery life with daily use8 days
operating temperature10 to 40°C