General Notes:
IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529. The compressor is protected against solid
foreign objects of 12 mm diameter and greater such as a finger, and
against vertically falling water drops which may cause issues during a
normal operation.
Subject to technical modification without prior notice.
The device may not work if the temperature and voltage conditions are
different to those defined in the specifications.
Performance may vary with drugs such as suspensions or high
viscosity. See drug supplier’s data sheet for further details.
Troubleshooting
In case of any of the below problems occur during use, first check that
no other electrical device is within 30 cm. If the problem persists, please
refer to the table below.
The device does not switch on
Check that the power plug is properly fitted to the power outlet.
Make sure the power switch is in the on position (
).
The device switches on but does not nebulize
Make sure the vaporizer head is fitted in the nebulizer kit.
Make sure the air tube is not squashed or crooked.
Check the air filter for blockage and dirt. Replace if necessary.
Check that sufficient amount of medication has been put into the
nebulizer kit.
The device suddenly stops working during operation.
The thermal cut-out has shut the device down for one of the following
reasons:
- the device was working in an environment with temperatures higher
than 40°C;
- the ventilation slots were covered.
Do not attempt to repair the device. Do not open and/or tamper with the
device. No parts of the device are user serviceable. Return the device to
an authorized OMRON retail outlet or distributor.
Please report to the manufacturer and the competent authority of the
Member State in which you are established about any serious incident
that has occurred in relation to this device.