Detailed information about your system
9
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 15mg/dL of a laboratory method
when the glucose concentration is lower than
100mg/dL, and within 15% of a laboratory method
when the glucose concentration is 100mg/dL or higher.
Samples from 100 patients were tested using both the
OneTouchSelect Plus Flex
System and the YSI 2300
Glucose Analyzer laboratory instrument.
System Accuracy Results for Glucose
Concentrations <100mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5mg/dL Within ±10mg/dL Within ±15mg/dL
67.2%
(121/180)
93.3%
(168/180)
96.7%
(174/180)