System Accuracy Results for Glucose
Concentrations ≥100mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15%
63.3%
(266/420)
95.5%
(401/420)
99.8%
(419/420)
System Accuracy Results for Glucose
Concentrations between 26.6mg/dL and
482.0mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±15mg/dL or ±15%
98.8% (593/600)
NOTE: Where 26.6mg/dL represents the lowest glucose
reference value and 482.0mg/dL represents the highest
glucose reference value (YSI value).