Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouchSelect Plus
Flex
System compares well with a laboratory method.
# of Subjects # of Tests Slope Intercept
(mg/dL)
100 600 0.98 -0.13
95% CI Slope 95% CI
Intercept
(mg/dL)
Std. Error
(S
y.x
)
(mg/dL)
R
2
0.97 to 0.98 -1.48 to 1.21 9.02 0.99