118
Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouchSelect Plus Flex
Meter.
Target
Glucose
(mg/dL)
Mean
Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient
of Variation
(%)
40 46.21 1.66 3.58
65 70.39 1.89 2.69
120 123.56 2.65 2.15
200 193.22 3.54 1.83
350 350.02 7.41 2.12
Results show that the greatest variability observed
between test strips when tested with blood is 1.89mg/dL
SD or less at glucose levels less than 100mg/dL, or 2.15%
CV or less at glucose levels at 100mg/dL or above.