Detailed information about your system
9
System Accuracy Results for Glucose
Concentrations ≥5.55mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15%
64.7% (268/414) 92.8% (384/414) 98.8% (409/414)
System Accuracy Results for Glucose
Concentrations between 2.0mmol/L and
28.2mmol/L
Within ±0.83mmol/L or ±15%
99.2% (595/600)
NOTE: Where 2.0mmol/L represents the lowest glucose
reference value and 28.2mmol/L represents the highest
glucose reference value (YSI value).