9 Detailed information about
your system
Comparing meter results to laboratory
results
Results obtained from the OneTouchUltra Plus Flex
Meter and laboratory tests are reported in plasma-
equivalent units. However, your meter result may differ
from your lab result due to normal variation. A result
from your OneTouchUltra Plus Flex
Meter is considered
accurate when it is within 0.83mmol/L of a laboratory
method when the glucose concentration is lower than
5.55mmol/L and within 15% of a laboratory method
when the glucose concentration is 5.55mmol/L or
higher.