112
Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouchUltra Plus Flex
Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 1.9 0.05 2.46
5.0 5.5 0.10 1.91
7.2 7.7 0.16 2.11
11.1 12.3 0.24 1.92
19.4 21.2 0.45 2.13
Results show that the greatest variability observed
between test strips when tested with blood is 1.89 mg/dL
SD or less at glucose levels less than 100 mg/dL, or 2.13%
CV or less at glucose levels at 100 mg/dL or above.