94
Troubleshooting your system
8
What it means
Meter is too hot (above 44°C)
to perform a test.
What to do
Move the meter and test strips to a cooler area. Insert a
new test strip when the meter and test strips are within
the operating range (6-44°C). If you do not get another
HI .t message, you can proceed with testing.
If this message continues to appear, contact Customer
Service. Contact OneTouch
®
Customer Service at
HK: (852) 2735 8262. MY: 1800 88 3338. SG:
+65 64157282.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OneTouch Ultra Plus Flex and is the answer not in the manual?

OneTouch Ultra Plus Flex Specifications

General IconGeneral
Display TypeLCD
ConnectivityBluetooth
Test Time5 seconds
Sample Size0.4 μL
Measurement Range20 to 600 mg/dL
Memory500 tests
Automatic ShutoffYes (after 2 minutes of inactivity)
Operating Conditions10% to 90% RH