TROUBLESHOOTING AND DETAILED INFORMATION ABOUT YOUR SYSTEM
Detailed information about your system
Technical specications
Reported result range 1.1–33.3 mmol/L
Calibration Plasma-equivalent
Sample Fresh capillary whole blood
Test time 5 seconds
Assay method Glucose oxidase biosensor
Meter power source One 3.0 Volt CR 2032 lithium battery or
equivalent
Unit of measure mmol/L
Memory 500 test results
Automatic shuto About 2 minutes after last action
Size 95 x 65 x 25 mm
Weight Approximately 58 grams, with battery
Operating ranges Temperature: 10–44°C
Relative humidity: 10–90%
Altitude: up to 3048 metres
Haematocrit: 30–55%
Battery ratings 1 x 3.0 V d.c., 60 mA
(1 x CR 2032 battery or equivalent)
direct current
System Accuracy
Diabetes experts have suggested that glucose meters should agree within 0.83 mmol/L of a laboratory method when the glucose concentration
is lower than 4.2 mmol/L, and within 20% of a laboratory method when the glucose concentration is 4.2 mmol/L or higher. Samples from
100 diabetic patients at 1 clinical center were tested using both the OneTouch® Vita™ System and the YSI Model 2300 Glucose Analyser
(laboratory test).