TROUBLESHOOTING AND DETAILED INFORMATION ABOUT YOUR SYSTEM
The meter has detected a problem
with the test strip. Possible causes are test
strip damage or an incompletely lled
conrmation window.
Repeat the test with a new test strip.
See Applying blood and reading results in
Section 2 or Section 5, Control solution
testing for instructions on properly lling
the conrmation window. Contact Customer
Care on 0800 121 200 (UK) or 1800 535 676
(Ireland).
Error message could be caused either by a
used/damaged test strip or a problem with
the meter.
Repeat the test with a new test strip; see
Applying blood and reading results in Section 2
or Section 5, Control solution testing.
Contact Customer Care on 0800 121 (UK)
or 1800 355 676 (ireland).
The sample was applied before the meter
was ready. If this message continues to appear
contact Customer Care on 0800 121 200 (UK)
or 1800 535 676 (Ireland)
Repeat the test with a new test strip. Apply
a blood or control solution sample only after
APPLY SAMPLE or APPLY CONTROL appears
on the display. If this message continues to
appear, contact Customer Care on 0800 121
(UK) or 1800 355 676 (ireland).
One of the following may apply:
You may have tested in an environment near
the low end of the system’s operating range.
or,
The test strip may have been
damaged or moved during testing.
or,
The sample was improperly applied.
or,
There may be a problem with the meter.
Make sure you are testing within the operating
range (10–44°C) and repeat the test with a
new test strip; see Applying blood and reading
results in Section 2 or Section 5, Control
solution testing. If this message continues to
appear contact Customer Care on 0800 121
200 (UK) or 1800 535 676 (Ireland).
WARNING
RETEST WITH A
NEW STRIP
ERROR 2
METER OR STRIP
PROBLEM
WARNING
RETEST WITH A
NEW STRIP
ERROR 3
METER WAS NOT
READY
WARNING
SEE USER’S HANDBOOK
ERROR 4
STRIP PROBLEM
WARNING
RETEST WITH A
NEW STRIP
ERROR 5
STRIP PROBLEM OR
SAMPLE TOO SMALL
Meter battery is low but still has enough
power to perform a test.
Once the LOW BATTERY/REPLACE BATTERY
SOON message is displayed, this symbol will
continue to appear until you replace the
battery. Test results will still be accurate, but
replace the battery as soon as possible; see
Changing the battery in Section 6.
Meter battery is low but still has enough
power to perform a test.
This message will appear when you turn
the meter on and there is enough power for
a minimum of 100 more tests. Press
to
continue, but replace the battery as soon as
possible; see Changing the battery in Section 6.
WARNING
SEE USER’S HANDBOOK
LOW BATTERY
REPLACE BATTERY
SOON
11 /OCT/08 09:30
AFTER MEAL
mmol/L
7. 7