TROUBLESHOOTING AND DETAILED INFORMATION ABOUT YOUR SYSTEM
Electrical and safety standards
This meter complies with CISPR 11: 2003, Class B (Radiated Only). Emissions of the energy used are low and not likely to cause interference
in nearby electronic equipment. The meter has been tested for immunity to Level 3 electrostatic discharge as specied in IEC 61000-4-2.
This meter has been tested for immunity to radio frequency interference over the frequency range 80 MHz to 2.5 GHz at 3 V/m as specied
in IEC 61000-4-3. Degree of protection rating: IP31
Guarantee
LifeScan guarantees that the OneTouch® Vita™ Meter will be free of defects in material and workmanship for three years, valid from the
date of purchase. The guarantee extends only to the original purchaser and is not transferable.
Precision
Within Run Precision (100 venous blood
tests per glucose level)
Total Precision (200 control solution tests
per glucose level)
Target Glucose
(mmol/L)
Mean Glucose
(mmol/L)
Standard
deviation
(mmol/L)
Coecient of
variation (%)
2.2 2.2 0.1 3.5
5.6 5.7 0.1 1.7
7.2 7.7 0.1 1.9
11.1 12.7 0.2 1.3
16.7 19.6 0.3 1.5
Glucose Levels
(mmol/L)
Mean Glucose
(mmol/L)
Standard
deviation
(mmol/L)
Coecient of
variation (%)
LOW 2.2 0.1 3.3
MID 6.1 0.1 1.8
HIGH 19.6 0.3 1.5
Results show that the greatest variability
observed between test strips when tested
with blood is 3.5% or less.