EasyManua.ls Logo

Optech Basic Series - Page 8

Default Icon
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ENGLISH 25
Figure 3
E
2.3 c) Now looking only through the left eye and using the left eyepiece (13 - ref. fig. 1),
adjust the focus on the specimen using just the dioptric compensation ring (14 - ref. fig. 1).
In this way, both eyes will view the specimen together in perfect focus.
2.4 Condenser centring (7 - ref. fig. 1) (B4 and B5 models only)
2.4 a) Selecting the 4x objective, close the condenser diaphragm (19 - ref. fig. 1)
2.4 b) Lower the condenser (8 - ref. fig. 1) using the condenser control knob (9 - ref. fig. 1)
until, observing through the eyepieces (13 - ref. fig. 1),
there is a sharp image of the internal edges
of the condenser diaphragm (19 - ref. fig. 1).
2.4 c) If the internal edges are not in the centre of the viewing
field, centre the condenser (8 - ref. fig. 1) by adjusting
the condenser centring screws (17 - ref. fig. 1).
2.5 Using the condenser diaphragm.
Adjusting the condenser diaphragm (19 - ref. fig. 1)
ensures that the specimen image displays the best balance
between resolution, luminosity, contrast and depth of field.
We suggest opening the diaphragm to approximately 70-80% of it’s maximum
aperture (see fig. 3). If the objective is changed then the condenser diaphragm
setting must be re-adjusted accordingly.
3. CARE AND MAINTENANCE
3.1 Cleaning
ATTENTION: always unplug the microscope before cleaning.
•To remove any dust on the lenses, use a soft bristle brush or lens cleaning tissue.
•To remove fingerprints or greasy marks from the lenses, use lens cleaning
tissue or a soft cloth soaked in a solvent such as xylene, benzene or ethyl
alcohol.The same solvents can also be used to clean oil immersion objectives
(16 - ref. fig. 1) NOTE: please remember to remove oil from these objectives
every time you finish using them.
•To clean the external, painted surfaces of the microscope,
use a gentle detergent and a soft cloth.
•We recommend that you always keep the microscope protected from dust
after use, a dust cover is provided for this purpose.
3.2 Please do not:
Remove the eyepieces (13 - ref. fig. 1) from the head, after fitting; dust ingress
could dirty prisms and internal lenses, reducing optical performance.
Use strong solvents to clean the external painted parts of the microscope;
they may damage the lacquer coating and plastic components.