OPF-P Series
User Manual / Using Scale
1
9/13
Quantity Check Alarm
1. Press QUANTITY CHECK key to enter quantity check setting
2. Select ON or OFF by pressing ZERO key and press M+ to go to next function
3. If OFF is selected, press M+ and the system exits back to weighing mode else
4. Input the lower quantity limit when the 2nd line ‘L’ flashes and press M+ to go
to next function
5. Input the upper quantity limit when the 3rd line ‘H’ flashes and press M+ to go
back to counting mode.
Sample Setting
1. Number setting (when the unit weight is unknown)
1. Place the # of samples on the platform and press CE key and the 2nd line
will show 0.
2. Set the number of samples using the numerical keypad
3. Press SAMPLE key; The display then shows CAL-- on the 3rd line.
2. Unit weight setting (when the unit weight is known)
1. Press CE key to cancel previous settings
2. Press UNIT WEIGHT key
Accumulation
1. Press M+ key in counting mode to enter into accumulation mode
2. Press UNIT WEIGHT key to save and go back to normal counting mode
3. The number of counts will be accumulated by repeating steps 1 and 2 (99 max)
4. Press CE key to clear the number of counts