OPF-P Series
User Manual / Using Scale
1
12/13
Communication Address
Selectable from 1 to 25
Enter the user setting mode, and press M+ until you see “SCALE?”.
Press ZERO or TARE key to select the value
Press M+ key to confirm the revised value or press the UNIT WEIGHT key to
cancel the changes.
BPS (baud rate)
Selectable between 2400, 4800 and 9600 (factory default)
Enter the user setting mode, and press M+ until you see “BAUD?”.
Press ZERO or TARE key to select the value
Press M+ key to confirm the revised value or press the UNIT WEIGHT key to
cancel the changes.
Transmission Method
Selectable between the following:
Continuous transmission
One stable transmission
Transmission as requested
Enter the user setting mode, and press M+ until you see “SENT?”.
Press ZERO or TARE key to select the value
Press M+ key to confirm the revised value or press the UNIT WEIGHT key to
cancel the changes.
RS232 Communication Protocols
If your scale comes with this option, the RS232 UART frame details are as follow:
10 bits
All ASCII code
Bit 1 : Start bit
Bit 2 - 9 : Data bits (8)
Bit 10 : Stop bit
RS232 9 pins
Pin 2 : TXD (send)
Pin 3 : RXD (receive)
Pin 5 : GND (ground)
Print label format
add: 1 : sub-add
n/w: 0.0001 : Net weight
u/w: 0.0000 : Unit weight
Pcs: 1 : Quantity