31
2. Remove the batteries or adapter before cleaning
the device.
Cleaning
Disinfect the mesh after each day's final treatment.
1. Remove the mouthpiece or mask from the
nebulizer and clean it.
2. After use, remove any remaining medication.
3. Add 6 ml of clean water (30-60 degrees Celsius)
to the medication container (2).
4. Turn on the device to remove all residues from
the mesh for 1 to 2 minutes.
5. If there are any traces of medical solutions or
impurities remaining on the gold-colored contact
points, use medical cotton swabs to remove
them.
6. Wipe the cap of the container with distilled water
and gently wipe the device.
Never wipe the mesh with sharp objects or touch it directly.
Also, do not use paper tissues, as this may interfere with the
normal functioning of the mesh.