35
Problem
Possible cause
The
nebulizer
produces
little or no
aerosol.
1. Too little medication in the vein.
2. The device is not held upright.
3. The medicine is not suitable for
nebulization.
4. The metal parts in the medicine
container (2) are dirty.
5. The batteries are not fully charged.
The output is
too low
1. The batteries are empty. Replace the
batteries or connect the adapter and try
again.
2. There are air bubbles in the
medication container that prevent the
medication from reaching the gauze.
Check this and remove the air bubbles.
3. Particles on the mesh hinder the
output. To remove this, fill the nebulizer
with 2 to 3 drops of vinegar and 3 to 6 ml
of water and nebulize this substance
completely. Do not inhale it and be sure
to clean and disinfect the device.
4. The mesh is worn.
Which
medications
are suitable
for
inhalation?
Only a doctor can advise which
medication is best to use to treat your
complaints. With a nebulizer you can
nebulize medication with a viscosity of
less than 3.
Some of the
inhalation
fluid remains
This is normal and happens for technical
reasons. Stop inhaling when you hear a