OrienTek T40 - Troubleshooting Common Issues; Fiber Alignment and Diameter Issues; Splice Parameter and Gas Issues

OrienTek T40
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
7.8
7.8
7.8
7.8 Heater
Heater
Heater
Heater Operation
Operation
Operation
Operation
Phenomenon Reason Resolve method
Fiber Not Aligned
Dust in the
V-groove
Clean
V-groove.
Fiber Diameter
Too Small
Discharge power is
inappropriate.
Do Arc Power
test and adjust
the current.
Splice parameter is
inappropriate.
Increase
discharge time or
stuff length.
Black Line
Splice parameter is
inappropriate.
Adjust the
current and the
discharge time.
Gas The fiber end is
unqualified.
Check cleaver.
The current too
small or discharge
time too short.
Adjust the
current and the
discharge time.
Fiber Asunder The stuff length is
too small.
Do an overlap
test.
The current is too
strong
Reduce the
current value and
the discharge
time.
Fiber Diameter
too big
The stuff length is
too big.
Do a overlap test.
The current is too
strong or the
discharge time too
long.
Adjust the current
value and the
discharge time.