27
Chapter
Chapter
Chapter
Chapter 8
8
8
8 Maintenance
Maintenance
Maintenance
Maintenance
8.1
8.1
8.1
8.1 Cleaning
Cleaning
Cleaning
Cleaning
Keep the V-groove, electrode, microscope clean. The shield
should be closed when the splicer is not used.
8.1.1
8.1.1
8.1.1
8.1.1 Cleaning
Cleaning
Cleaning
Cleaning V-groove
V-groove
V-groove
V-groove
If there is dust on the V-groove, the bare fiber alignment
operation may produce a big splice loss. Clean the V-groove in
the following steps.
(1) Open the shield .
(2) Cleave a bare fiber and push the tip of this fiber from the
V-groove
’
s one end to the other end.
(3) If the above method cannot clean the V-groove, you can
use an alcohol-soaked cotton swab to brush the groove to
get rid of the dust.
8.1.2
8.1.2
8.1.2
8.1.2 Clean
Clean
Clean
Clean ing
ing
ing
ing Electrodes
Electrodes
Electrodes
Electrodes
(1) If the electrodes are polluted because of long-term use ,
wipe the tip of electrodes lightly with a cotton stick dipped
in acetone or a 3 mm wide and 50 mm
l
ong grinding paper.
Pay attention to protecting the electrodes' tip and its
position during cleaning. F inally stabilize electrodes, see
Clear
Clear
Clear
Clear and
and
and
and Stabilize
Stabilize
Stabilize
Stabilize Electrodes
Electrodes
Electrodes
Electrodes
(Section D.4) for detail.
(2) Replacing electrodes is a complex techni cal job , which can
only be done by the manufacturer when the splicer is sent
back for an overhauling.
8.1.3
8.1.3
8.1.3
8.1.3 Cleaning
Cleaning
Cleaning
Cleaning the
the
the
the Objective
Objective
Objective
Objective Lens
Lens
Lens
Lens
The surface of the objective lens must be kept clean. Dust on