Operation
Daily routine
1. Check that the red lamp is not lit. If the lamp is on the temperature has been below 2 degrees
or above 8 degrees. For more information on alarm see the section on alarm. (figure 7)
2. Check the defrost water container (the large bottle), empty as necessary.
In the case of alarm
If an alarm is triggered the red alarm lamp lights and an audible alarm sounds for 120 minutes. In the
event of an alarm the user must "acknowledge the alarm". This is carried out as follows.
1. Press ESC. This shuts off the acoustic signal.
2. Press the menu button.
3. Use the arrow to move two steps to the right to select the "ALARM" icon.
4. Press OK to display information about the latest alarm.
5. The red lamp goes out if the temperature is within the alarm limit 2-8 °C.
Description of the display
Below is an overview of the buttons on the display. For further information about the settings go to
page 9.
MENU button
To access the menu and
its settings. Press MENU
once and then use the
right/left button to
navigate the menu.
ESC button
To cancel or alternatively
to return to the normal
display or shut off the
sound in the event of an
alarm.
ON/OFF
Alarm lamp
Lights red in the
event of an alarm
OK button
Confirm the desired
command.