6
180625
Installing the medication and pharmacy cabinet
1. The cabinet is set to an operating temperature of +5 °C which is measured in 50 ml of fluid.
The alarm range is +2-+8 °C. When the power cable is plugged into the wall socket the display
will show the current temperature and PF.
2. Press briefly on the button at the far left to reset PF (Button No. 2).
3. Wait 4 hours, or until the cabinet displays the correct temperature. The cabinet is now ready
for use.
4. Note that the refrigerator shelves are designed for a maximum load of 12 kg.
5. Note that goods must not be stored on the bottom of the refrigerator.
6. Note that goods must not be loaded outside the shelf and/or against the back of the
refrigerator.
7. To ensure optimal circulation in the refrigerator there should be an airflow between the goods,
in other words between the packaging.
8. The power supply to the medication cabinet is 230 V ~50 Hz, 95 W.
9. The refrigeration unit refrigerant is R134a.
10. The refrigerator has a service class of 6.
11. Do not place heavy objects on top of the medication cabinet.
12. For refrigerators with a volume of 140 litres and above: To keep the refrigerator stable mount
the adjustable feet supplied. Adjust the feet by turning them.
13. The pharmacy cabinet can be provided with an external alarm cable. Connect to NC when
installing.
14. The length of the alarm cable is 2 metres.
15. If the power cable is damaged it must be replaced by authorised personnel or by the
manufacturer.
16. The length of the power cable is 3.5 metres.
17. The refrigerator is factory calibrated.