6
of the AFO.
6. To relieve contact between the medial strut and the apex of the medial
malleolus, grind the edge of the strut according to the shape shown (figure 8).
- you may grind up to 1 cm into the strut
- stay at least 7 cm from the footplate
GENERAL REMARKS AND ADDITIONAL ADVICE:
1. Toshortenthefootplate,cutorgrindotheappropriatelength.Becertainto
butheedgestoeliminatesharpedges.Tolengthenthe
footplate, use foot insert of choice, and fixate it to the foot plate.
2. To adjust the biomechanics of the foot, the use of the correct insert is advised.
3. Contraindications: Do not use the AFO in case of excessive ankle instability
or in case excessive deformities to the foot where pressure relief cannot be
solvedbyadditionalpadding.ItiscontraindicatedtousetheAFOincaseof
fluctuation oedema and in case of severe spasticity.
4. TheAFOcannotbereshapedbyheating.Heatingwillcausedelaminatingand
negatively alter the dynamics of the orthosis.
5. The right footwear is critically important for proper function of the AFO. The
following criteria should be met when selecting the right
footwear to use with this AFO:
• Firmheelcounter-forpropercontroloftherearfoot
• Firmshank-totakestressotheAFO
• LacedorVelcrofastening-foreasierdonninganddong,andtoallow
adjustable compressive support over the mid-foot
• TheshoeshavetobewideenoughtoallowproperttingoftheAFO
inside the shoe
• Removableinsole-toallowspacefortheAFOfootplatewith
modifications
• Approximately6/9“(1,7cm)heelheight-asastartingpointtocontrol
the knee extension
For optimal comfort and function as well as proper customization of the ÖSSUR
AFO: Dynamic, the use of an orthopaedic insole is recommended.
WARNING
With the use of this type of device, it is possible to receive various types of skin
irritationandorpressurepoints.Ifyouexperienceanyunusualdiscomfortwhile
using this product, consult your health care provider immediately.
This product has been designed and tested based on single patient usage and is
notrecommendedformultiplepatientuse.Ifanyproblemsoccurwiththeuseof
this product, for example, if unusual movement or product wear is detected in a
structural part of the product, immediately discontinue use of the device and
contactyourmedicalprofessional.CustomizationbeyondthatspecifiedinthisIFU
is not recommended.