101
* Battery current is measured according to IEC 60118-0:1983/AMD1:1994 §7.11, IEC 60118-0:2015 §7.7 and ANSI
S3.22:2014 §6.13 after a settling time of minimum 3 minutes.
** Real usage battery life is shown as an estimated interval based on mixed use cases with variable amplification
settings and variable input levels, incl. direct stereo streaming from a TV (25% of the time) and streaming from
a mobile phone (6% of the time).
0 dB SPL ref. 20 μPa Oticon Real 1 Oticon Real 2 & 3
Battery Consumption*
Typical 2.0 mA 2.0 mA
Quiescent 1.9 mA 1.9 mA
Expected Battery Life
(battery size 312 -
IEC PR41)**
Hours 50-55 hrs 50-55 hrs
Latency 8.4 ms 8.4 ms
Maximum Induction Coil
Sensitivity
Measured output at 1 mA/m 75 dB SPL 75 dB SPL
Measured output at 10 mA/m 95 dB SPL 95 dB SPL
Measured output at 31.6 mA/m 105 dB SPL 105 dB SPL