EasyManuals Logo

OTT Pluvio User Manual

Default Icon
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
from contact of housing
123123
Connection of
weighing cell
CONTACT
black red
Shield
white
CONTACT
20 mA +
20 mA -
green
3
123456
123456
CONTACT
Status (10 Hz)
Resolution 0.01 mm
black
1
2
green /
pink
Connection
for +12 V DC
3
Supply current source -
violet
grey
black
pink
grey
CONTACT
red /
blue
20 mA +
Supply current source +
20 mA -
yellow
green
red
blue
Supply GND
Supply +12 VDC
n. c.
n. c.
+12 V
n. c.
white
brown
red/blue
Status
Pluvio precipitation
yellow
green/pink
violet
grey
pink
blue
red
black
brown
green
white
2
1
1
2
+12 VDC (supply)
GND (supply)
8…24V(supply current source)
GND (supply current source)
Connection
for sensors/outputs
Pluvio type 1000 with status output
}
PWR
Contact (0.1 mm; 10 Hz)
Status (10 Hz)
Status (10 Hz)
Contact (0.1 mm; 10 Hz)
20 mA (current source; 0.1 mm)
20 mA + (current source; status)
20 mA (current source; status)
20 mA + (current source; 0.1 mm)
ATTENTION
Resolution: 0.1 mm per pulse, output frequency 10 Hz symmetric
1.0 mm per pulse, output frequency 1 Hz symmetric
Range: 0 to 30 mm precipitation per minute