7 Maintenance
INFORMATION
The product may be exposed to increased loads by the patient.
► Shorten the maintenance intervals according to the expected loads.
The manufacturer requires an inspection of the product every 6months to
verify functionality and check for wear at a minimum.
Spare parts are listed in the section "Components/design".
17AD1*
A service kit can be ordered for product maintenance.
Replacing the compression spring
► Replace the compression spring with a new spring when there are signs
of wear and tear, but no later than after 12months.
► Lubricate the orthotic ankle joints, using only 633F7 special lubricant.
Before handing the orthosis over to the patient, secure all screw connec
tions with Loctite241.
17AF2*
► The orthotic ankle joints are maintenance-free.
If you have any doubts, please contact the manufacturer.
8 Disposal
Dispose of the product in accordance with national regulations.
9 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in accord
ance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregarding the
information in this document, particularly due to improper use or unauthor
ised modification of the product.
9.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medic
al devices. The CE declaration of conformity can be downloaded from the
manufacturer's website.