4. Check the three pressures of the Intensity legs button on the Controller.
If the pressure rates are within certain ranges, the Air bag and the Pump unit are
OK. If the rates are low, check the pressure of the Pump unit. If the Pump unit is
OK, there are something wrong with the Air bag or the Electromagnetic valves.
Caution
Be careful not to bend hoses while checking.