Panasonic MCO-170AICUVHL - User Manual

Panasonic MCO-170AICUVHL
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
MCO-170AICL
MCO-170AICUVL
MCO-170AICUVHL
Please read the operating instructions carefully before using this product, and save the operating
instructions for future use.
See page 86 for all model numbers.
This product is for research use only.
Not for clinical diagnosis or treatment of humans or animals.
CO
2
Incubator
Operatin
g
Instructions

Questions and Answers

  • J
    Jeffrey RodriguezAug 20, 2025
    Why Panasonic MCO-170AICUVHL chamber temperature does not match the set value?
    • D
      Dr. Mary Bates DDSAug 20, 2025
      If the chamber temperature on your Panasonic Incubator isn't matching the set value, check if the ambient temperature is less than 5 ?C different from the set chamber temperature. Also, make sure the outer door is closed and the inner door isn't left open. Lastly, ensure the incubator isn't near equipment that generates electromagnetic waves.
  • D
    davidmitchellAug 23, 2025
    What to do if Panasonic MCO-170AICUVHL CO gas density does not match the set value?
    • L
      lleblancAug 23, 2025
      If the CO gas density in your Panasonic Incubator isn't matching the set value, verify that the secondary pressure for the gas regulator is at the specified value of 0.03 MPa(G) (0.3 kgf/cm2(G), 4.3 psi(G)). Check for any blockages in the gas tube and ensure the duct is securely attached to the 4 points hooks. Also, confirm that the fan is properly attached and pushed all the way to the motor shaft. Finally, ensure the incubator is not operating near equipment that generates electromagnetic waves.
  • K
    kristina04Aug 26, 2025
    What to do if Panasonic MCO-170AICUVHL Incubator CO gas density is slow to recover?
    • T
      taylorscottAug 26, 2025
      If the CO gas density is slow to recover in your Panasonic Incubator, and the CO gas pressure is normal, the HEPA filter may be clogged and you should contact our sales representative or agent. Also, check if there is little gas remaining in the CO gas cylinder and ensure the secondary pressure for the gas regulator is at the specified value of 0.03 MPa(G) (0.3 kgf/cm2(G), 4.3 psi(G)). Check for blockages in the gas tube and ensure the duct is securely attached to the 4 points hooks. Confirm that the fan is properly attached and pushed all the way to the motor shaft.
  • A
    Alison PearsonAug 28, 2025
    What to do if Panasonic Incubator is consuming a large quantity of CO gas?
    • Q
      qsaundersAug 28, 2025
      If your Panasonic Incubator is consuming a large quantity of CO gas, check if the outer and inner doors are being frequently opened and closed. Inspect connectors for gas leaks due to deterioration of the gas tube or pinhole leaks; consider replacing the gas tube annually. Also, verify if the packing seal for the inner door is defective and if the access hole is open.
  • D
    Dr. Jeffrey WilcoxAug 30, 2025
    What to do if Panasonic MCO-170AICUVHL chamber humidity does not rise?
    • S
      Shawn WhiteAug 30, 2025
      If the chamber humidity in your Panasonic Incubator isn't rising, ensure there is enough sterile distilled water in the humidifying pan.
  • J
    jerry91Sep 2, 2025
    Why normal cultures are not possible, and the CO gas density is suspect in Panasonic MCO-170AICUVHL Incubator?
    • R
      Richard StokesSep 2, 2025
      If normal cultures are not possible and the CO gas density is suspect in your Panasonic Incubator, check if the ambient air environment around the incubator is normal and if there is a source of polluted gas nearby.
  • C
    Christopher Sparks IIISep 4, 2025
    What to do if Panasonic MCO-170AICUVHL does not operate at all?
    • K
      Kelly OneillSep 4, 2025
      If your Panasonic Incubator isn't operating, check if it's properly plugged in and if there's a power outage or a tripped circuit breaker. Also, verify that the removal power supply cord is connected to the port on the lower right side of the cabinet.
  • R
    Regina KelleySep 7, 2025
    Why Panasonic Incubator outer door does not open?
    • S
      Stephen ParkerSep 7, 2025
      If the outer door of your Panasonic Incubator doesn't open, it may be because the electric lock is engaged when the power switch is OFF. Turn the power switch ON or use the accessory unlock key to override the electric lock. Also, the outer door is electrically locked during decontamination.
  • L
    lporterSep 9, 2025
    Why CO gas is not being injected in Panasonic MCO-170AICUVHL?
    • S
      Shawn ThompsonSep 10, 2025
      The CO control method for the Panasonic Incubator is the ON-OFF method. CO gas is intermittently injected as the gas density in the chamber approaches the set value. Injections may be stopped for periods of approximately 15 seconds, but that is not an error. The gas is not injected until the temperature of the CO sensor becomes stable enough approx. 1 hour, after turning ON the power switch or recovering from power failure.
  • U
    uvazquezSep 12, 2025
    Why Panasonic Incubator UV lamp lights when the outer door is open?
    • L
      Leslie GarciaSep 12, 2025
      If the UV lamp lights when the outer door of your Panasonic Incubator is open, check if something is pushing the door switch.

Summary

Precautions for Safe Operation

General Safety Symbols and Warnings

Explains hazard symbols (WARNING, CAUTION) and fundamental safety advice.

Electrical and Installation Safety

Details precautions for electrical safety, grounding, and proper installation environment.

Safe Handling and Usage Practices

Covers safe handling procedures, maintenance, storage, and disposal advice.

Installation Guide

Site and Environmental Requirements

Specifies the necessary environmental conditions and location characteristics for installation.

Physical Installation Steps

Outlines the procedure for physically installing the unit, including leveling and grounding.

CO2 Gas Cylinder Connection

Details the safe and correct procedure for connecting the CO2 gas cylinder.

Before Commencing Operation

For Better Cultivation

Correct Operation Procedure

Basic Operation on LCD Touch Panel

Basic Parameters Configuration

Setting Temperature and CO2 Levels

Details how to input numerical values and set chamber temperature and CO2 density.

Alarm Parameters Setup

Temperature and CO2 Alarm Settings

Explains how to configure automatic temperature and CO2 density alarm thresholds.

Alarm Delay and Notification Settings

Details settings for alarm delay, ring back function, and remote alarm output.

UV Lamp Parameters

H2 O2 Decontamination Procedure

H2 O2 Decontamination Process

Details the step-by-step procedure for chamber decontamination using H2O2.

Routine Maintenance

Alarms, Safety, and Self-Diagnosis

Troubleshooting Common Issues

Panasonic MCO-170AICUVHL Specifications

General IconGeneral
TypeCO2 Incubator
Temperature Control Range5°C above ambient to 50°C
Temperature Stability±0.1°C
CO2 Control Range0 to 20%
CO2 Stability±0.1%
Capacity170L
Internal Volume170 L
Chamber Volume170 L
UV SterilizationYes
Temperature Range5°C above ambient to 50°C
Temperature Uniformity±0.3°C
CO2 Accuracy±0.15%
CO2 Sensor TypeInfrared CO2 Sensor
Humidity Range95% RH at 37°C
Shelves3
Power SupplyAC 230V, 50/60Hz
Humidity95% RH at 37°C