The filter on the front plate of your PARI BOY S / PARI BOY SX compressor
must be replaced every 200 operating hours in normal operating conditions,
or at least once a year.
Also please check the filter at regular intervals (after every 10 to 12 inhala-
tions). If it is soiled (grey or brown colour) or blocked, replace it. If the filter
has become damp, it must also be replaced with a new filter
1)
.
Do not try to clean the filter and use it again.
Only use original PARI filters; otherwise your PARI compressor may be dam-
aged or treatment may be inadequate.
To change the filter, proceed as follows (see figure D):
Unscrew the filter holder (7) from your PARI BOY S / PARI BOY SX coun-
ter-clockwise using a suitable coin.
Pull the old filter (11) out of the filter holder and fit the new filter in its place.
Insert the filter holder with the new filter in your PARI BOY S /
PARI BOY SX and screw it in firmly, turning clockwise with a coin.
This product falls within the scope of the European Council Directive on
Waste Electrical and Electronic Equipment (WEEE
2)
) and is included in prod-
uct category 8: Medical devices for the purposes thereof. Accordingly, this
product must not be disposed of with domestic waste. The disposal regula-
tions prevailing in the respective member countries must be observed (e.g.,
disposal by local authorities or dealers).
Materials recycling helps to reduce the consumption of raw materials and pro-
tect the environment.
1) Part. no. 041B4852, available from your dealer or PARI GmbH.
2) Directive 2002/96/EC of the EUROPEAN PARLIAMENT AND THE EUROPEAN
COUNCIL of January 27, 2003 on waste electrical and electronic equipment.
7 Filter change
8Disposal