EasyManua.ls Logo

Pari PRONEB MAX - Maintenance and Repair; Compressor Maintenance and Filter Replacement

Pari PRONEB MAX
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Questions? 1.800.FAST.NEB (327.8632)
|
PARI.com/PRONEBMAX
EN-6
G. MAINTENANCE OF YOUR
The filter on the front of the PRONEB
®
max Compressor
was designed to be monitored without removing it. This
makes it easy to check without much work. When the
filter is new, it is bright white in color. With use, the
filter will turn grey, indicating the need for replacement.
The filter must be replaced immediately if it gets wet or
becomes clogged. A good rule to follow is to replace
your filter every 6 months - at the same time as your
PARI reusable nebulizer.
For filter replacement, obtain the replacement filter
set part #130F1002P2 which includes the filter and
filter cap (Fig E). Pull the old filter and cap out of the
compressor and discard (Fig F).
Once the old filter and cap are removed, tilt the
compressor so that the face of the unit is facing the
floor. Gently tap on the back of the compressor to
remove any unwanted materials from the filter area.
Place the compressor on a stable surface and using
clean hands, push a new filter with cap into the filter
receptacle.
NOTE: DO NOT ATTEMPT TO WASH AND
REUSE AIR FILTERS. DO NOT USE SUBSTITUTE
FILTER MATERIAL AS COMPRESSOR DAMAGE
MAY RESULT.
COMPRESSOR
Use a damp cloth to wipe the exterior of your PRONEB
®
max housing and tubing.
CAUTION
Never spray fluid into the ventilation slots on your
PRONEB
®
max. Fluid could cause damage to the electrical
& compressor parts and could lead to malfunction.
In the event fluid enters the unit, please contact PARI
Patient Connection at 1.800.FAST.NEB (327.8632).
MAINTENANCE & REPAIR
Never operate your PRONEB
®
max if it is improperly or
incompletely assembled or damaged. Use only original
PARI parts and attachments authorized by PARI for the
PRONEB
®
max Aerosol Delivery System. Use of other
parts and attachments may damage the unit or cause
improper treatment. Do not open or disassemble the
compressor unit at any time.
NOTE: THERE ARE NO SERVICEABLE PARTS
INSIDE THE PRONEB
®
MAX AEROSOL DELIVERY
SYSTEM. SERVICE BY ANYONE VOIDS
WARRANTY.
E F
PRONEB
®
MAX
PRONEB
®
max REPLACEMENT ITEMS
Part No.
LC Sprint
®
Reusable Nebulizer 023F35
LC PLUS
®
Reusable Nebulizer 022F81
Wing Tip
®
Tubing 044E0050
PRONEB
®
max Replacement Filter 2pk 130F1002P2

Related product manuals