EasyManuals Logo

Pentax EC-380LKp User Manual

Pentax EC-380LKp
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
- 25 -
(2) REPROCESSING SYSTEM
To reprocess a PENTAX single-channel videoscope, first an enzymatic detergent and then a high-level disinfectant
or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and scope components
(air/water valve, suction valve, etc.). Exposure times of detergent and disinfectant/sterilant must be strictly adhered
to as recommended by each respective manufacuturer.
Please note that all solution entrance ports and flow pathways are illustrated above.
CAUTION:
BEFORE IMMERSING:
a) The ‘Red’ ETO gas sterilization venting cap must be taken OFF.
b) The PVE Soaking cap should be securely ON the electrical contacts.
NOTE:
Prior to exposure of all internal channels to an enzymatic detergent and high-level disinfectant (or steri-
lant), all brushable channels should be manually cleaned with cleaning brushes.
EG-290Kp/EC-380Kp
Water Feeding Channel
Rubber Inlet Seal
Channel Inlet
A/W Instrument Channel Cleaning Adapter OF-B115
Air/Water Channel
Cleaning Adapter OF-G17
PVE Soaking Cap
(ATTACH)
ETO Cap
(REMOVE)
Air Feeding Tube
Water Feeding Tube
Suction Tube
Suction Nipple
Luer-Slip Syringe
with
Cleaning/Disinfectiing
Solution
Luer-Slip Syringe
with
Cleaning/Disinfectiing
Solution
Air Feeding Channel
Instrument Channel
Air/Water Nozzle

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pentax EC-380LKp and is the answer not in the manual?

Pentax EC-380LKp Specifications

General IconGeneral
BrandPentax
ModelEC-380LKp
CategoryMedical Equipment
LanguageEnglish

Related product manuals