- 4 -
1-3. VIDEO PROCESSOR
CAUTION:
Please refer to the instruction supplied with the processor.
NOTE:
Do not use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle.
Although the cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and
flow of air and water during the endoscopic procedure. Both the PENTAX water bottle cap and bottle
(container) are identified by their appropriate model designation. Ensure that an OS-H4 cap is used only
with the OS-H4 water container/bottle.
Do not overtighten the bottle cap. Overtighting can cause the bottle cap to break.
WARNING:
The lifetime of the lamp in EPK-i processor is 500 hours. Prior to use, check the lamp life meter on the rear
panel to ensure the lamp life is less than 500 hours. After 500 hours of use, the image quality will
deteriorate. Excessive use of the lamp beyond 500 hours could break the lamp inside the equipment
resulting in damage to the equipment.
WARNING:
The lifetime of the lamp in EPK-i processor is 500 hours. Prior to use, check the lamp life meter on the front
panel to ensure the lamp life is less than 500 hours. After 500 hours of use, the image quality will
deteriorate. Excessive use of the lamp beyond 500 hours could break the lamp inside the equipment
resulting in damage to the equipment.
NOTE:
The lamp life could be affected by frequency of use. In which case, the lamp life might become shorter than
its respective rated hours (EPK-i: 500 hours, EPK-1000 and EPK-100p: 400 hours)
WATER BOTTLE
LAMP SWITCH
AIR PUMP SWITCH
LCD
POWER SWITCH
INTERFACE SOCKET